Back to Search Start Over

Performance standards in positron emission tomography.

Authors :
Karp JS
Daube-Witherspoon ME
Hoffman EJ
Lewellen TK
Links JM
Wong WH
Hichwa RD
Casey ME
Colsher JG
Hitchens RE
Source :
Journal of nuclear medicine : official publication, Society of Nuclear Medicine [J Nucl Med] 1991 Dec; Vol. 32 (12), pp. 2342-50.
Publication Year :
1991

Abstract

A standard set of performance measurements is proposed for use with positron emission tomographs. This set of measurements has been developed jointly by the Computer and Instrumentation Council of the Society of Nuclear Medicine and the National Electrical Manufacturers Association. The measurements include tests of spatial resolution, scatter fraction, sensitivity, count rate losses and randoms, uniformity, scatter correction, attenuation correction, and count rate linearity correction.

Details

Language :
English
ISSN :
0161-5505
Volume :
32
Issue :
12
Database :
MEDLINE
Journal :
Journal of nuclear medicine : official publication, Society of Nuclear Medicine
Publication Type :
Academic Journal
Accession number :
1744726