Back to Search Start Over

Incomplete seating of the liner with the Trident acetabular system: a cause for concern?

Authors :
Langdown AJ
Pickard RJ
Hobbs CM
Clarke HJ
Dalton DJ
Grover ML
Source :
The Journal of bone and joint surgery. British volume [J Bone Joint Surg Br] 2007 Mar; Vol. 89 (3), pp. 291-5.
Publication Year :
2007

Abstract

We reviewed the initial post-operative radiographs of the Trident acetabulum and identified a problem with seating of the metal-backed ceramic liner. We identified 117 hips in 113 patients who had undergone primary total hip replacement using the Trident shell with a metal-backed alumina liner. Of these, 19 (16.4%) were noted to have incomplete seating of the liner, as judged by plain anteroposterior and lateral radiographs. One case of complete liner dissociation necessitating early revision was not included in the prevalence figures. One mis-seated liner was revised in the early post-operative period and two that were initially incompletely seated were found on follow-up radiographs to have become correctly seated. There may be technical issues with regard to the implanting of this prosthesis of which surgeons should be aware. However, there is the distinct possibility that the Trident shell deforms upon implantation, thereby preventing complete seating of the liner.

Details

Language :
English
ISSN :
2044-5377
Volume :
89
Issue :
3
Database :
MEDLINE
Journal :
The Journal of bone and joint surgery. British volume
Publication Type :
Academic Journal
Accession number :
17356136
Full Text :
https://doi.org/10.1302/0301-620X.89B3.18473