Incomplete seating of the liner with the Trident acetabular system: a cause for concern?
- Source :
-
The Journal of bone and joint surgery. British volume [J Bone Joint Surg Br] 2007 Mar; Vol. 89 (3), pp. 291-5. - Publication Year :
- 2007
-
Abstract
- We reviewed the initial post-operative radiographs of the Trident acetabulum and identified a problem with seating of the metal-backed ceramic liner. We identified 117 hips in 113 patients who had undergone primary total hip replacement using the Trident shell with a metal-backed alumina liner. Of these, 19 (16.4%) were noted to have incomplete seating of the liner, as judged by plain anteroposterior and lateral radiographs. One case of complete liner dissociation necessitating early revision was not included in the prevalence figures. One mis-seated liner was revised in the early post-operative period and two that were initially incompletely seated were found on follow-up radiographs to have become correctly seated. There may be technical issues with regard to the implanting of this prosthesis of which surgeons should be aware. However, there is the distinct possibility that the Trident shell deforms upon implantation, thereby preventing complete seating of the liner.
- Subjects :
- Acetabulum surgery
Adult
Aged
Arthroplasty, Replacement, Hip methods
Cohort Studies
Female
Hip Joint surgery
Humans
Joint Diseases surgery
Male
Middle Aged
Postoperative Care methods
Prosthesis Design
Prosthesis Failure
Radiography
Reoperation
Acetabulum diagnostic imaging
Arthroplasty, Replacement, Hip instrumentation
Hip Joint diagnostic imaging
Hip Prosthesis
Subjects
Details
- Language :
- English
- ISSN :
- 2044-5377
- Volume :
- 89
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- The Journal of bone and joint surgery. British volume
- Publication Type :
- Academic Journal
- Accession number :
- 17356136
- Full Text :
- https://doi.org/10.1302/0301-620X.89B3.18473