Back to Search
Start Over
Hybrid polymer/thin film impedance system for label free monitoring of cells.
- Source :
-
Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference [Conf Proc IEEE Eng Med Biol Soc] 2004; Vol. 2004, pp. 2561-4. - Publication Year :
- 2004
-
Abstract
- Impedance measurements (capacitive cytometry) have been used to perform label-free analysis of cells. Two devices were constructed using a simple liquid photopolymerization technique known as muFluidicTectonics. This platform has made it possible to rapidly fabricate with great ease and without the use of traditional MEMS technology a cell impedance measurement system. Measurements were made using not only a single 1 kHz frequency but also a frequency sweep from 50 Hz to 20 kHz. Single frequency measurements on SP2 mouse cells showed show that there is a strong, linear relationship between the DNA content of individual cells and their dielectric (or capacitance) response to a 1 kHz field. Frequency sweeps were conducted on SF9 cells as means to perform diagnostic measurements of the device when different number of cells are present.
Details
- Language :
- English
- ISSN :
- 1557-170X
- Volume :
- 2004
- Database :
- MEDLINE
- Journal :
- Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference
- Publication Type :
- Academic Journal
- Accession number :
- 17270796
- Full Text :
- https://doi.org/10.1109/IEMBS.2004.1403736