[FE-analysis of surface stresses for the tribological system in total hip prostheses].
- Source :
-
Biomedizinische Technik. Biomedical engineering [Biomed Tech (Berl)] 2006 Dec; Vol. 51 (5-6), pp. 367-70. - Publication Year :
- 2006
-
Abstract
- The implantation of a total hip prosthesis is an operation which is performed frequently due to advanced hip joint damage both in humans and in veterinary medicine in dogs. The long-term result of a hip prosthesis is mainly determined by aseptic loosening of the prosthesis; among other causes, abrasion particles of the tribological pairing are responsible for the loosening. For the analysis of the surface stresses with different tribological pairings, a finite element model was generated which was based on the CAD data of a commercial total hip prosthesis. After transmission of a physiological force in the components of the three tribological pairings ceramic/polyethylene, ceramic/ceramic and metal/polyethylene, stresses were calculated. Stresses in the ceramic/ceramic tribological pairings were conspicuously higher than in the other material pairings. In the future adapted prostheses have to be developed that ensure optimal friction and absorption characteristics of the components.
- Subjects :
- Arthroplasty, Replacement, Hip methods
Computer Simulation
Finite Element Analysis
Friction
Humans
Lubrication
Prosthesis Failure
Stress, Mechanical
Surface Properties
Arthroplasty, Replacement, Hip instrumentation
Equipment Failure Analysis methods
Hip Joint physiopathology
Hip Joint surgery
Hip Prosthesis
Models, Biological
Subjects
Details
- Language :
- German
- ISSN :
- 0013-5585
- Volume :
- 51
- Issue :
- 5-6
- Database :
- MEDLINE
- Journal :
- Biomedizinische Technik. Biomedical engineering
- Publication Type :
- Academic Journal
- Accession number :
- 17155874
- Full Text :
- https://doi.org/10.1515/BMT.2006.071