Diagnosis of early cardiac transplant rejection by fall in evoked T wave amplitude measured using an externalized QT driven rate responsive pacemaker.

Authors :
Grace AA
Newell SA
Cary NR
Scott JP
Large SR
Wallwork J
Schofield PM
Source :
Pacing and clinical electrophysiology : PACE [Pacing Clin Electrophysiol] 1991 Jun; Vol. 14 (6), pp. 1024-31.
Publication Year :
1991

Abstract

Reliable diagnosis of cardiac allograft rejection is at present only possible using endomyocardial biopsy. We have serially measured epicardial evoked T wave amplitude during ventricular pacing with an externalized QT driven rate responsive pacemaker telemetered to a TP2 analyzer in 13 patients (12 males) followed for 19 (14-26) days after transplantation. A total of 228 records were analyzed. Rejection was defined on endomyocardial biopsy. On 17 of the 31 occasions on which biopsy was performed during the study, specimens showed significant (moderate) rejection. In 11 patients the initial biopsy proven rejection episode was associated with a significant fall in the evoked T wave amplitude from 1.3 (0.7-2.3) mV to 0.6 (0.5-1.8) mV (P less than 0.005), which began 2 (1-4) days earlier. One patient with uncontrolled diabetes mellitus had no change in evoked T wave amplitude during rejection. The evoked T wave amplitude did not fall in the absence of histologic rejection. These results suggest a noninvasive method for detecting cardiac rejection, which appears both sensitive (92%) and specific (100%) in the first rejection episodes.

Details

Language :
English
ISSN :
0147-8389
Volume :
14
Issue :
6
Database :
MEDLINE
Journal :
Pacing and clinical electrophysiology : PACE
Publication Type :
Academic Journal
Accession number :
1715063
Full Text :
https://doi.org/10.1111/j.1540-8159.1991.tb04153.x