Back to Search Start Over

Setup accuracy of stereoscopic X-ray positioning with automated correction for rotational errors in patients treated with conformal arc radiotherapy for prostate cancer.

Authors :
Soete G
Verellen D
Tournel K
Storme G
Source :
Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology [Radiother Oncol] 2006 Sep; Vol. 80 (3), pp. 371-3. Date of Electronic Publication: 2006 Aug 17.
Publication Year :
2006

Abstract

We evaluated setup accuracy of NovalisBody stereoscopic X-ray positioning with automated correction for rotational errors with the Robotics Tilt Module in patients treated with conformal arc radiotherapy for prostate cancer. The correction of rotational errors was shown to reduce random and systematic errors in all directions. (NovalisBody and Robotics Tilt Module are products of BrainLAB A.G., Heimstetten, Germany).

Details

Language :
English
ISSN :
0167-8140
Volume :
80
Issue :
3
Database :
MEDLINE
Journal :
Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology
Publication Type :
Academic Journal
Accession number :
16914219
Full Text :
https://doi.org/10.1016/j.radonc.2006.07.001