[Determination of the error of calculation of the effective energy of combined X-ray beam].

Authors :
Petrushanskiĭ MG
Kornev EA
Pishchukhin AM
Source :
Meditsinskaia tekhnika [Med Tekh] 2006 May-Jun (3), pp. 20-2.
Publication Year :
2006

Abstract

A formula for assessing the error of calculation of the effective energy of combined X-ray beam is derived. Maximal error of calculation of the effective energy of combined X-ray beam is found to be (4% for anode voltage of 40-150 kV and aluminum filter thickness of 2-45 mm.