Back to Search Start Over

Near fatal arrhythmia induction by an implantable cardioverter defibrillator.

Authors :
Wollmann CG
Göttker U
Böcker D
Gradaus R
Source :
Journal of cardiovascular electrophysiology [J Cardiovasc Electrophysiol] 2006 Sep; Vol. 17 (9), pp. 1026-8. Date of Electronic Publication: 2006 Jun 09.
Publication Year :
2006

Abstract

Sudden death among implantable cardioverter defibrillator (ICD) recipients is well known. The underlying mechanism remains unclear in most cases. One possible mechanism is an accidentally induced arrhythmia by the ICD itself. This report describes for the first time a case of a life-threatening ventricular arrhythmia induced by a specific algorithm for ventricular arrhythmia reconfirmation before shock delivery in a fourth-generation ICD.

Details

Language :
English
ISSN :
1540-8167
Volume :
17
Issue :
9
Database :
MEDLINE
Journal :
Journal of cardiovascular electrophysiology
Publication Type :
Academic Journal
Accession number :
16764706
Full Text :
https://doi.org/10.1111/j.1540-8167.2006.00532.x