Back to Search Start Over

An improved dual channel PERCA instrument for atmospheric measurements of peroxy radicals.

Authors :
Green TJ
Reeves CE
Fleming ZL
Brough N
Rickard AR
Bandy BJ
Monks PS
Penkett SA
Source :
Journal of environmental monitoring : JEM [J Environ Monit] 2006 May; Vol. 8 (5), pp. 530-6. Date of Electronic Publication: 2006 Apr 21.
Publication Year :
2006

Abstract

This paper describes a new dual-channel PEroxy RadiCal Amplification (PERCA) instrument, which has been designed to improve the time resolution and signal to noise and to reduce the interference caused by variations in ambient ozone concentrations. The instrument was run at the Weybourne Atmospheric Observatory (WAO), North Norfolk, during WAOWEX (Weybourne Atmospheric Observatory Winter Experiment) in January/February 2002 and INSPECTRO (Influence of clouds on the spectral actinic flux in the lower troposphere) in September 2002. The performance of the instrument is assessed and compared to that of a single channel instrument. In particular, it is shown how the precision is greatly improved in fluctuating background ozone conditions. In addition the improved time response of the instrument allows changes in peroxy radical concentrations to be related to rapid changes in nitric oxide concentrations and the ozone photolysis frequency, j(O(1)D).

Details

Language :
English
ISSN :
1464-0325
Volume :
8
Issue :
5
Database :
MEDLINE
Journal :
Journal of environmental monitoring : JEM
Publication Type :
Academic Journal
Accession number :
16688354
Full Text :
https://doi.org/10.1039/b514630e