In vivo estimation of the flow-driven adiabatic inversion efficiency for continuous arterial spin labeling: a method using phase contrast magnetic resonance angiography.
- Source :
-
Magnetic resonance in medicine [Magn Reson Med] 2006 Jun; Vol. 55 (6), pp. 1291-7. - Publication Year :
- 2006
-
Abstract
- The accurate quantification of perfusion with arterial spin labeling (ASL) requires consideration of a number of factors, including the efficiency of the inversion and control pulses used for spin labeling. In this study the effects of spin velocity on continuous ASL efficiency when using the amplitude modulated control strategy were investigated using simulations of the Bloch equations. The inversion efficiency was determined in vivo by combining the simulations with phase-contrast velocity mapping data acquired at the level of the tagging plane. Using this novel method, an average inversion efficiency of 69% was calculated for a group of 28 subjects, in good agreement with experimental data reported previously. There was, however, a large range in inversion efficiency measured across the subject group (50-76%), indicating that the velocity dependence of the amplitude modulated control efficiency may introduce additional variability into the perfusion calculations if not properly taken into account.<br /> (Copyright 2006 Wiley-Liss, Inc.)
- Subjects :
- Arteries anatomy & histology
Humans
Information Storage and Retrieval methods
Reproducibility of Results
Sensitivity and Specificity
Algorithms
Arteries physiology
Blood Flow Velocity physiology
Image Enhancement methods
Image Interpretation, Computer-Assisted methods
Magnetic Resonance Angiography methods
Subjects
Details
- Language :
- English
- ISSN :
- 0740-3194
- Volume :
- 55
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Magnetic resonance in medicine
- Publication Type :
- Academic Journal
- Accession number :
- 16673361
- Full Text :
- https://doi.org/10.1002/mrm.20864