Validation of a new high-performance liquid chromatography assay for nizatidine.

Authors :
Yusuf A
Al Dgither S
Hammami MM
Source :
Therapeutic drug monitoring [Ther Drug Monit] 2006 Apr; Vol. 28 (2), pp. 232-6.
Publication Year :
2006

Abstract

An expedient high-performance liquid chromatography (HPLC) assay for nizatidine measurement in human plasma was developed and validated. After deproteinization of 200 microL of plasma by filtration, nizatidine and 4-amino-antipyrine (internal standard) were separated (capacity ratio 3.0 and 6.63, respectively) on Nova-Pak C18 cartridge at room temperature (RT), and detected spectrophotometrically at 320 nm. The mobile phase, 0.02 mol/L disodium hydrogen phosphate, acetonitrile, methanol, and triethylamine (80:10:10:0.05 vol/vol), was delivered at 1.5 mL/min. Calibration curves were linear (r2 > or = 0.999) in the range 0.02 to 5 microg/mL, detection and quantification limits were 0.01 and 0.02 microg/mL, respectively, intra-run and inter-run coefficients of variation were < or = 3.5% and < or = 4.2%, respectively, and recovery was >90%. Nizatidine was stable for at least 4 hours at RT, 12 weeks at -20 degrees C, and 3 freeze-thaw cycles in plasma; 16 hours at RT and 48 hours at -20 degrees C in deproteinized plasma; and 6 hours at RT and 3 weeks at -20 degrees C in water.

Details

Language :
English
ISSN :
0163-4356
Volume :
28
Issue :
2
Database :
MEDLINE
Journal :
Therapeutic drug monitoring
Publication Type :
Academic Journal
Accession number :
16628136
Full Text :
https://doi.org/10.1097/01.ftd.0000189902.08684.90