Back to Search Start Over

Comparison of electrospray ionization mass spectrometry and evaporative light scattering detections for the determination of Poloxamer 188 in itraconazole injectable formulation.

Authors :
Nair LM
Konkel J
Thomas M
Koberda M
Source :
Journal of pharmaceutical and biomedical analysis [J Pharm Biomed Anal] 2006 Jun 07; Vol. 41 (3), pp. 725-30. Date of Electronic Publication: 2006 Feb 03.
Publication Year :
2006

Abstract

A high performance liquid chromatography (HPLC) method for Poloxamer 188 using size-exclusion chromatography (SEC) was developed and two different detection mechanisms, evaporative light scattering (ELSD) and electrospray ionization mass spectrometry (ESI-MS), were compared for their quantification capabilities in itraconazole formulation. Both detection techniques coupled with SEC separation were highly effective for the determination of Poloxamer 188, which is difficult to analyze by other common HPLC methods. As expected, ESI-MS detection provided sensitivity and selectivity superior to ELSD. But since the analyte is an excipient in the formulation, high sensitivity was not required and ELSD's simplicity and ruggedness made it more appropriate for routine analysis of this formulation.

Details

Language :
English
ISSN :
0731-7085
Volume :
41
Issue :
3
Database :
MEDLINE
Journal :
Journal of pharmaceutical and biomedical analysis
Publication Type :
Academic Journal
Accession number :
16458475
Full Text :
https://doi.org/10.1016/j.jpba.2005.12.028