Back to Search Start Over

Wavelet-based low-delay ECG compression algorithm for continuous ECG transmission.

Authors :
Kim BS
Yoo SK
Lee MH
Source :
IEEE transactions on information technology in biomedicine : a publication of the IEEE Engineering in Medicine and Biology Society [IEEE Trans Inf Technol Biomed] 2006 Jan; Vol. 10 (1), pp. 77-83.
Publication Year :
2006

Abstract

The delay performance of compression algorithms is particularly important when time-critical data transmission is required. In this paper, we propose a wavelet-based electrocardiogram (ECG) compression algorithm with a low delay property for instantaneous, continuous ECG transmission suitable for telecardiology applications over a wireless network. The proposed algorithm reduces the frame size as much as possible to achieve a low delay, while maintaining reconstructed signal quality. To attain both low delay and high quality, it employs waveform partitioning, adaptive frame size adjustment, wavelet compression, flexible bit allocation, and header compression. The performances of the proposed algorithm in terms of reconstructed signal quality, processing delay, and error resilience were evaluated using the Massachusetts Institute of Technology University and Beth Israel Hospital (MIT-BIH) and Creighton University Ventricular Tachyarrhythmia (CU) databases and a code division multiple access-based simulation model with mobile channel noise.

Details

Language :
English
ISSN :
1089-7771
Volume :
10
Issue :
1
Database :
MEDLINE
Journal :
IEEE transactions on information technology in biomedicine : a publication of the IEEE Engineering in Medicine and Biology Society
Publication Type :
Academic Journal
Accession number :
16445252
Full Text :
https://doi.org/10.1109/titb.2005.856854