Back to Search Start Over

Spectroscopic imaging of the brain with phased-array coils at 3.0 T.

Authors :
Xu D
Chen AP
Cunningham C
Osorio JA
Nelson SJ
Vigneron DB
Source :
Magnetic resonance imaging [Magn Reson Imaging] 2006 Jan; Vol. 24 (1), pp. 69-74. Date of Electronic Publication: 2005 Dec 19.
Publication Year :
2006

Abstract

The goal of this study was to develop and evaluate high-resolution magnetic resonance spectroscopic imaging (MRSI) utilizing the gains in signal-to-noise ratio (SNR) provided by combining higher magnetic field with high-sensitivity phased-array (PA) coils. We investigated the maximum improvement in spatial resolution as small as 0.09 cm(3) for brain MRSI while maintaining adequate SNR and acquisition time. The use of low peak power, dual-band spectral-spatial pulses was also investigated for application to 3 T MRSI of the brain using the body coil for radiofrequency excitation and PA coils for signal reception.

Details

Language :
English
ISSN :
0730-725X
Volume :
24
Issue :
1
Database :
MEDLINE
Journal :
Magnetic resonance imaging
Publication Type :
Academic Journal
Accession number :
16410180