Back to Search Start Over

Development of micro-PIC as a time-resolved X-ray area detector.

Authors :
Takeda A
Uekusa H
Kubo H
Miuchi K
Nagayoshi T
Ohashi Y
Okada Y
Orito R
Takada A
Tanimori T
Source :
Journal of synchrotron radiation [J Synchrotron Radiat] 2005 Nov; Vol. 12 (Pt 6), pp. 820-5. Date of Electronic Publication: 2005 Oct 18.
Publication Year :
2005

Abstract

The application of a two-dimensional micro-pixel gas chamber (micro-PIC) to X-ray diffraction studies, and its performance, are reported. micro-PIC has a 10 cm x 10 cm detection area, and a fast-readout system for real-time X-ray imaging has been developed. Using the timing of each incoming X-ray measured by micro-PIC, continuous rotation photograph measurements were carried out for a 400 microm-diameter spherical organic crystal of Ylid (C(11)H(10)O(2)S), and then diffraction spots were successfully obtained within 2theta of 49 degrees. As a result, good internal agreement factors of 3.7% and 7.0% were confirmed among each of the symmetrically equivalent reflections for exposure times of 3700 s and 98 s, respectively.

Details

Language :
English
ISSN :
0909-0495
Volume :
12
Issue :
Pt 6
Database :
MEDLINE
Journal :
Journal of synchrotron radiation
Publication Type :
Academic Journal
Accession number :
16239754
Full Text :
https://doi.org/10.1107/S0909049505026713