Back to Search Start Over

Titrimetric method for determination of O-desulfated heparin in physiological samples using protamine-sensitive membrane electrode as endpoint detector.

Authors :
Buchanan SA
Kennedy TP
MacArthur RB
Meyerhoff ME
Source :
Analytical biochemistry [Anal Biochem] 2005 Nov 15; Vol. 346 (2), pp. 241-5. Date of Electronic Publication: 2005 Sep 15.
Publication Year :
2005

Abstract

O-Desulfated heparin (ODSH) is a promising new anti-inflammatory agent for the prevention of reperfusion injury following myocardial infarction or stroke. This partially desulfated heparin derivative has less anticoagulant activity than unfractionated heparin but retains the inherent anti-inflammatory properties of heparin. Thus, ODSH could be administered at the high doses needed to achieve desired anti-inflammatory function without risk of hemorrhage. However, given the very low anticoagulant activity of this species, traditional methods for heparin determination in clinical samples might not be well suited for ODSH measurements. In this article, a novel titrimetric method for detection of ODSH in buffer and plasma is described using a protamine-sensitive polymer membrane electrode as the detector. Titrations of ODSH with the heparin antagonist protamine yield sharp endpoints with sensitivity to ODSH in the micrograms per milliliter range for plasma samples. The stoichiometry for protamine interaction with ODSH is determined to average 1.39 microg protamine/microg ODSH in plasma. This technology is further applied to a toxicokinetic study of ODSH in an animal model, demonstrating the ability to detect the changes in ODSH concentrations in biological samples.

Details

Language :
English
ISSN :
0003-2697
Volume :
346
Issue :
2
Database :
MEDLINE
Journal :
Analytical biochemistry
Publication Type :
Academic Journal
Accession number :
16213460
Full Text :
https://doi.org/10.1016/j.ab.2005.08.035