Measurement of dose-area product with GafChromic XR Type R film.

Authors :
Thomas G
Li Y
Chu RY
Cheung JY
Maqbool F
Rabe F
Burns GS
Source :
Journal of applied clinical medical physics [J Appl Clin Med Phys] 2005 Summer; Vol. 6 (3), pp. 122-32. Date of Electronic Publication: 2005 Aug 12.
Publication Year :
2005

Abstract

Many of the newer X-ray machines are equipped with electronic means to provide dose-area product (DAP) information. For machines without that ability, an alternative method is to record radiation on a film that can handle a large amount of cumulative exposure. The use of GafChromic XR Type R film was investigated for this purpose by placing it at the X-ray tube assembly to record the radiation in interventional radiological procedures. Dose-area product was determined with a reflective densitometer and then with a flatbed scanner. Precisions were demonstrated to be 5% and 2%, respectively. In a comparison with the machine-recorded DAP, a regression analysis showed the validity of both techniques for values less than 1200 Gy-cm2.