Characterization of a new MOSFET detector configuration for in vivo skin dosimetry.
- Source :
-
Medical physics [Med Phys] 2005 Jun; Vol. 32 (6), pp. 1571-8. - Publication Year :
- 2005
-
Abstract
- The dose released to the patient skin during a radiotherapy treatment is important when the skin is an organ at risk, or on the contrary, is included in the target volume. Since most treatment planning programs do not predict dose within several millimeters of the body surface, it is important to have a method to verify the skin dose for the patient who is undergoing radiotherapy. A special type of metal oxide semiconductors field-effect transistors (MOSFET) was developed to perform in vivo skin dosimetry for radiotherapy treatments. Water-equivalent depth (WED), both manufacturing and sensor reproducibility, dependence on both field size and angulation of the sensor were investigated using 6 MV photon beams. Patient skin dosimetries were performed during 6 MV total body irradiations (TBI). The resulting WEDs ranged from 0.04 and 0.15 mm (0.09 mm on average). The reproducibility of the sensor response, for doses of 50 cGy, was within +/-2% (maximum deviation) and improves with increasing sensitivity or dose level. As to the manufacturing reproducibility, it was found to be +/-0.055 mm. No WED dependence on the field size was verified, but possible variations of this quantity with the field size could be hidden by the assessment uncertainty. The angular dependence, for both phantom-surface and in-air setups, when referred to the mean response, is within +/-27% until 80 degree rotations. The results of the performed patient skin dosimetries showed that, normally, our TBI setup was suitable to give skin the prescribed dose, but, for some cases, interventions were necessary: as a consequence the TBI setup was corrected. The water-equivalent depth is, on average, less than the thinnest thermoluminescent dosimeters (TLD). In addition, when compared with TLDs, the skin MOSFETs have significant advantages, like immediate both readout and reuse, as well as the permanent storage of dose. These sensors are also waterproof. The in vivo dosimetries performed prove the importance of verifying the dose to the skin of the patient undergoing radiotherapy.
- Subjects :
- Calibration
Humans
Monte Carlo Method
Phantoms, Imaging
Photons
Radiation Dosage
Radiotherapy Dosage
Radiotherapy Planning, Computer-Assisted
Radiotherapy, Conformal
Reproducibility of Results
Sensitivity and Specificity
Software
Ultrasonics
Whole-Body Irradiation
Radiometry instrumentation
Radiometry methods
Skin pathology
Skin radiation effects
Subjects
Details
- Language :
- English
- ISSN :
- 0094-2405
- Volume :
- 32
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Medical physics
- Publication Type :
- Academic Journal
- Accession number :
- 16013716
- Full Text :
- https://doi.org/10.1118/1.1924328