Back to Search
Start Over
Evaluation of the flexural strength of carbon fiber-, quartz fiber-, and glass fiber-based posts.
- Source :
-
Journal of endodontics [J Endod] 2005 Mar; Vol. 31 (3), pp. 209-11. - Publication Year :
- 2005
-
Abstract
- This study investigated the flexural strength of eight fiber posts (one carbon fiber, one carbon/quartz fiber, one opaque quartz fiber, two translucent quartz fiber, and three glass fiber posts). Eighty fiber posts were used and divided into eight groups (n = 10): G1: C-POST (Bisco); G2: AESTHETI-POST (Bisco); G3: AESTHETI-PLUS (Bisco); G4: LIGHT-POST (Bisco); G5: D.T. LIGHT-POST (Bisco); G6: PARAPOST WHITE (Coltene); G7: FIBERKOR (Pentron); G8: REFORPOST (Angelus). All of the samples were tested using the three-point bending test. The averages obtained were submitted to the ANOVA and to Tukey's test (p < 0.05). The mean values (MPa) of the groups AESTHETI-POST-carbon/quartz fiber post (Bisco) and AESTHETI-PLUS-quartz fiber post (Bisco) were statistically similar and higher than the mean values of the other groups. The mean values of the groups C-POST-carbon fiber post (Bisco), LIGHT-POST-translucent quartz fiber post (Bisco), D.T. LIGHT-POST-double tapered translucent quartz fiber post (Bisco), PARAPOST WHITE-glass fiber post (Coltene) and FIBREKOR--glass fiber post (Pentron) were similar and higher than the group REFORPOST-glass fiber post (Angelus).
Details
- Language :
- English
- ISSN :
- 0099-2399
- Volume :
- 31
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Journal of endodontics
- Publication Type :
- Academic Journal
- Accession number :
- 15735472
- Full Text :
- https://doi.org/10.1097/01.don.0000137652.49748.0c