Reduction of RF penetration effects in high field imaging.

Authors :
Foo TK
Hayes CE
Kang YW
Source :
Magnetic resonance in medicine [Magn Reson Med] 1992 Feb; Vol. 23 (2), pp. 287-301.
Publication Year :
1992

Abstract

A method of correcting for the RF inhomogeneity in the body by dielectric loading of the coil-to-shield space in an RF resonator (coil and shield assembly) is described. The presence of the RF coil and RF shield have significant effects on RF homogeneity. Based on theoretical calculations, a method for adjusting the RF homogeneity by manipulating the axial propagation constant, kappa z, is proposed. This is accomplished by loading the coil-to-shield space with dielectric material of suitable relative permittivity so as to increase kappa z and decrease the radial propagation constant, kappa rho. In this manner, the radial wavelength (lambda rho = 2 pi/kappa rho) can be increased relative to the body dimensions, and the field amplitude variations in the axial plane minimized. Theoretical calculations indicate that a value of between 30 and 40 for the relative permittivity of the dielectric material in the coil-to-shield space would reduce the RF field inhomogeneity from +/- 15% to about +/- 3% over a central 30-cm-diameter region of a homogeneous 40-cm-diameter body at both 64 and 170 MHz. The theoretical model was verified in laboratory measurements of the [formula; see text] field generated in a test coil at 170 MHz which was scaled to correspond to a body at 64 MHz. However, the improved RF field homogeneity would be accompanied by increased RF power requirements and reduced coil sensitivity.

Details

Language :
English
ISSN :
0740-3194
Volume :
23
Issue :
2
Database :
MEDLINE
Journal :
Magnetic resonance in medicine
Publication Type :
Academic Journal
Accession number :
1549043
Full Text :
https://doi.org/10.1002/mrm.1910230209