Computational fluid dynamics analysis of a maglev centrifugal left ventricular assist device.

Authors :
Burgreen GW
Loree HM 2nd
Bourque K
Dague C
Poirier VL
Farrar D
Hampton E
Wu ZJ
Gempp TM
Schöb R
Source :
Artificial organs [Artif Organs] 2004 Oct; Vol. 28 (10), pp. 874-80.
Publication Year :
2004

Abstract

The fluid dynamics of the Thoratec HeartMate III (Thoratec Corp., Pleasanton, CA, U.S.A.) left ventricular assist device are analyzed over a range of physiological operating conditions. The HeartMate III is a centrifugal flow pump with a magnetically suspended rotor. The complete pump was analyzed using computational fluid dynamics (CFD) analysis and experimental particle imaging flow visualization (PIFV). A comparison of CFD predictions to experimental imaging shows good agreement. Both CFD and experimental PIFV confirmed well-behaved flow fields in the main components of the HeartMate III pump: inlet, volute, and outlet. The HeartMate III is shown to exhibit clean flow features and good surface washing across its entire operating range.