Back to Search Start Over

Poly(dimethylsiloxane) microchip capillary electrophoresis with electrochemical detection for rapid measurement of acetaminophen and its hydrolysate.

Authors :
He FY
Liu AL
Xia XH
Source :
Analytical and bioanalytical chemistry [Anal Bioanal Chem] 2004 Aug; Vol. 379 (7-8), pp. 1062-7. Date of Electronic Publication: 2004 Jun 18.
Publication Year :
2004

Abstract

Poly(dimethylsiloxane) microchip capillary electrophoresis with amperometric detection has been used for rapid separation and determination of acetaminophen and its hydrolysate, i.e. p-aminophenol. A Pt ultramicroelectrode with a diameter of 10 microm positioned at the outlet of the separation channel was used as a working electrode for amperometric detection. Factors influencing separation and detection were investigated and optimized. Results show that acetaminophen and p-aminophenol can be well separated within 35 s with RSD-1 (approximately 0.1 fmol) at S/ N=3. This method has been successfully applied to the detection of traces of p-aminophenol in paracetamol tablets.

Details

Language :
English
ISSN :
1618-2642
Volume :
379
Issue :
7-8
Database :
MEDLINE
Journal :
Analytical and bioanalytical chemistry
Publication Type :
Academic Journal
Accession number :
15221194
Full Text :
https://doi.org/10.1007/s00216-004-2680-4