Back to Search Start Over

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation.

Authors :
Linthout N
Verellen D
Van Acker S
Storme G
Source :
Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology [Radiother Oncol] 2004 May; Vol. 71 (2), pp. 235-41.
Publication Year :
2004

Abstract

A spreadsheet based program is presented to perform an independent Monitor Unit (MU) calculation verification for the Quality Assurance (QA) of Intensity Modulated Radiation Therapy (IMRT) using Dynamic MultiLeaf Collimation (DMLC). The computed dose value is compared to the planned dose by calculating the percent dose difference per Intensity Modulated Beam (IMB) and absolute dose difference per IMB. The proposed acceptability levels are +/-5.0% or +/-2.0 cGy for the percent dose difference per IMB and the absolute dose difference per IMB, respectively. For percent dose difference per treatment, an acceptability level of +/-2.0% is proposed. The presented program is considered adequate for checking the treatment plans calculated for IMRT treatments using DMLC as a part of the QA procedure.

Details

Language :
English
ISSN :
0167-8140
Volume :
71
Issue :
2
Database :
MEDLINE
Journal :
Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology
Publication Type :
Academic Journal
Accession number :
15110458
Full Text :