A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation.
- Source :
-
Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology [Radiother Oncol] 2004 May; Vol. 71 (2), pp. 235-41. - Publication Year :
- 2004
-
Abstract
- A spreadsheet based program is presented to perform an independent Monitor Unit (MU) calculation verification for the Quality Assurance (QA) of Intensity Modulated Radiation Therapy (IMRT) using Dynamic MultiLeaf Collimation (DMLC). The computed dose value is compared to the planned dose by calculating the percent dose difference per Intensity Modulated Beam (IMB) and absolute dose difference per IMB. The proposed acceptability levels are +/-5.0% or +/-2.0 cGy for the percent dose difference per IMB and the absolute dose difference per IMB, respectively. For percent dose difference per treatment, an acceptability level of +/-2.0% is proposed. The presented program is considered adequate for checking the treatment plans calculated for IMRT treatments using DMLC as a part of the QA procedure.
- Subjects :
- Dose-Response Relationship, Radiation
Humans
Particle Accelerators instrumentation
Phantoms, Imaging
Quality Control
Radiation Injuries prevention & control
Radiation Oncology
Radiotherapy Dosage
Radiotherapy Planning, Computer-Assisted methods
Radiotherapy, Conformal methods
Sensitivity and Specificity
Software
Radiation Monitoring instrumentation
Radiation Tolerance
Radiotherapy Planning, Computer-Assisted instrumentation
Radiotherapy, Conformal instrumentation
Subjects
Details
- Language :
- English
- ISSN :
- 0167-8140
- Volume :
- 71
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology
- Publication Type :
- Academic Journal
- Accession number :
- 15110458
- Full Text :
- https://doi.org/10.1016/j.radonc.2004.02.014