Back to Search Start Over

Determination of myo-inositol in biological samples by liquid chromatography-mass spectrometry.

Authors :
Perelló J
Isern B
Costa-Bauzá A
Grases F
Source :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences [J Chromatogr B Analyt Technol Biomed Life Sci] 2004 Apr 05; Vol. 802 (2), pp. 367-70.
Publication Year :
2004

Abstract

Due to the absence of HPLC methods to determine myo-inositol using mass detection and considering its sensitivity and selectivity, a high performance liquid chromatography-mass spectrometry method for the analysis of myo-inositol is described and applied to its direct determination in urine and saliva samples. Successful resolution of myo-inositol and its related substances was achieved with a stationary phase Aminex HPX-87C Column with milli-Q water as mobile phase and 5 mM ammonium acetate added post-column. The detector counted positive ions by monitoring m/z = 198, which corresponds to the myo-inositol adduct with ammonium cation. Urine and saliva samples were previously purified by passing through an anion-exchange resin. Concentrations as low as 138 and 461 microg/l in saliva and urine could be respectively quantified. Intra-day R.S.D. ranged from 0.83 to 1.02%, whereas inter-day R.S.D. was between 1.54 and 3.58%.

Details

Language :
English
ISSN :
1570-0232
Volume :
802
Issue :
2
Database :
MEDLINE
Journal :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences
Publication Type :
Academic Journal
Accession number :
15018800
Full Text :
https://doi.org/10.1016/j.jchromb.2003.12.005