Back to Search Start Over

Comparison of diamond-like carbon-coated stents versus uncoated stainless steel stents in coronary artery disease.

Authors :
Airoldi F
Colombo A
Tavano D
Stankovic G
Klugmann S
Paolillo V
Bonizzoni E
Briguori C
Carlino M
Montorfano M
Liistro F
Castelli A
Ferrari A
Sgura F
Di Mario C
Source :
The American journal of cardiology [Am J Cardiol] 2004 Feb 15; Vol. 93 (4), pp. 474-7.
Publication Year :
2004

Abstract

Stainless steel (SS) and carbon-coated (CC) stents were randomly compared in 347 patients (520 lesions). No differences were observed in in-hospital major adverse cardiac events: 2.8% in the CC group and 4.5% in the SS group (p = 0.286). The 6-month follow-up showed similar rates of binary restenosis (31.8% in the CC group vs 35.9% in the SS group; p = 0.448) and of cumulative major adverse cardiac events (30.5% in the CC group vs 32.7% in the SS group; p = 0.675). In unselected patients and lesions, carbon coating does not provide significant improvements over SS stents with the same design.

Details

Language :
English
ISSN :
0002-9149
Volume :
93
Issue :
4
Database :
MEDLINE
Journal :
The American journal of cardiology
Publication Type :
Academic Journal
Accession number :
14969628
Full Text :
https://doi.org/10.1016/j.amjcard.2003.10.048