Back to Search Start Over

Development of a FIA system with immobilized enzymes for specific post-column detection of purine bases and their nucleosides separated by HPLC column.

Authors :
Yao T
Matsumoto Y
Wasa T
Source :
Journal of biotechnology [J Biotechnol] 1990 Apr; Vol. 14 (1), pp. 89-97.
Publication Year :
1990

Abstract

A sensitive and highly selective method for the simultaneous determination of purine bases and their nucleosides is proposed. An amperometric flow-injection system with the two immobilized enzyme reactors (guanase immobilized reactor and purine nucleoside phosphorylase/xanthine oxidase co-immobilized reactor) is used as the specific post-column detection system of HPLC, to convert compounds separated by a reversed-phase. HPLC column to electroactive species (hydrogen peroxide and uric acid) which can be detected at a flow-through platinum electrode. The proposed detection system is specific for a group of purine bases and purine nucleosides and does not respond for purine nucleotides and pyrimidine bases. The linear determination ranges are from 10 pmol to 5 nmol for four purine bases (hypoxanthine, xanthine, guanine, and adenine) and four purine nucleosides (inosine, xanthosine, guanosine, and adenosine). The detection limits are 1.2-5.5 pmol.

Details

Language :
English
ISSN :
0168-1656
Volume :
14
Issue :
1
Database :
MEDLINE
Journal :
Journal of biotechnology
Publication Type :
Academic Journal
Accession number :
1367430
Full Text :
https://doi.org/10.1016/0168-1656(90)90021-3