Back to Search Start Over

Determination of huperzine A in formulated products by reversed-phase-liquid chromatography using diode array and electrospray ionization mass spectrometric detection.

Authors :
Yang QP
Kou XL
Fugal KB
McLaughlin JL
Source :
Phytomedicine : international journal of phytotherapy and phytopharmacology [Phytomedicine] 2003 Mar; Vol. 10 (2-3), pp. 200-5.
Publication Year :
2003

Abstract

A precise and selective high-performance liquid chromatographic (HPLC) method with diode-array detection for quantifying huperzine A in formulated products was developed and validated. A liquid chromatographic-mass spectrometric (LC/MS) procedure was devised to confirm the HPLC method. Huperzine A was dissolved in 1,2-dichloroethane, chromatographed on a YMCBasic C18 column, and detected at 308 nm. A gradient mobile phase of 10 mM ammonium acetate (pH = 3.5)--methanol was used. Identification was based on retention time, UV spectra and mass spectra by comparison with a commercial standard. The UV peak areas were used for quantitation of huperzine A content. The correlation coefficient (R2) of the calibration curve was 1 over the range 0.8-11.6 microg/ml. Overall recovery of huperzine A was 103.9% +/- 1.8 (mean +/- SD). Relative standard deviations for intra- and interday precision were < 2%.

Details

Language :
English
ISSN :
0944-7113
Volume :
10
Issue :
2-3
Database :
MEDLINE
Journal :
Phytomedicine : international journal of phytotherapy and phytopharmacology
Publication Type :
Academic Journal
Accession number :
12725577
Full Text :
https://doi.org/10.1078/094471103321659942