Stereotaxic assembly and procedures for simultaneous electrophysiological and MRI study of conscious rat.

Authors :
Khubchandani M
Mallick HN
Jagannathan NR
Mohan Kumar V
Source :
Magnetic resonance in medicine [Magn Reson Med] 2003 May; Vol. 49 (5), pp. 962-7.
Publication Year :
2003

Abstract

A stereotaxic restraining assembly was designed and developed for simultaneous electrophysiological recordings and functional MRI (fMRI) data acquisition from a conscious rat. The design of the nonmagnetic stereotaxic apparatus facilitated the restraining of head and body of the unanesthetized conscious animal during MRI experiments. The apparatus was made of Teflon and Perspex materials with an appropriate size and shape for a 4.7 T / 40 cm animal MRI scanner. Electrodes made from nonmagnetic silver wire were implanted on the skull for recording the electroencephalogram (EEG), the electro-oculogram (EOG), and the electromyogram (EMG), while polycarbonate screws were used for anchoring the electrode assembly. There were no major distortions or artifacts observed in the electrophysiological tracings and MR images. Electrophysiological recordings during fMRI acquisitions are useful to study different neurophysiological mechanisms of sleep and pathophysiology of seizure activity. Integration of electrophysiological recordings (with their good temporal resolution) and MRI (with its superior spatial resolution) is helpful in characterizing the functional state of different brain regions.<br /> (Copyright 2003 Wiley-Liss, Inc.)

Details

Language :
English
ISSN :
0740-3194
Volume :
49
Issue :
5
Database :
MEDLINE
Journal :
Magnetic resonance in medicine
Publication Type :
Academic Journal
Accession number :
12704780
Full Text :
https://doi.org/10.1002/mrm.10441