Back to Search Start Over

[A chemiluminescence flow sensor for vitamin B1].

Authors :
Song ZH
Huang JX
Wang R
Source :
Yao xue xue bao = Acta pharmaceutica Sinica [Yao Xue Xue Bao] 2001 Jun; Vol. 36 (6), pp. 467-9.
Publication Year :
2001

Abstract

Aim: To determine vitamin B1 in pharmaceutical preparations by chemiluminescence (CL) flow sensor.<br />Methods: When 200 microL of Na3PO4 was passed through an anion exchange column, K3Fe(CN)6 was eluted from the resin and then mixed with the vitamin B1 stream containing NaOH to product CL, by the fast oxidation reaction between vitamin B1 and K3Fe(CN)6 in alkaline solution.<br />Results: The CL emission intensity was correlated with the vitamin B1 concentration in the range 1.0 x 10(-5) to 1.0 x 10(-3) mol.L-1 and the detection limit was 8.0 x 10(-6) mol.L-1 (3 sigma). A complete analysis, including sampling and washing, can be performed within 2 min with a relative standard deviation of less than 4.0%.<br />Conclusion: The vitamin B1 flow sensor was stable for over 200 analyses and has been applied successfully to determination of vitamin B1 in pharmaceutical preparations.

Details

Language :
Chinese
ISSN :
0513-4870
Volume :
36
Issue :
6
Database :
MEDLINE
Journal :
Yao xue xue bao = Acta pharmaceutica Sinica
Publication Type :
Academic Journal
Accession number :
12585135