Right ventricular apexangiography for precise placement of transvenous electrode catheters.

Authors :
Stertzer SH
DePasquale NP
Bruno MS
Cohn LJ
Source :
Chest [Chest] 1976 Feb; Vol. 69 (2), pp. 222-3.
Publication Year :
1976

Abstract

A simple angiographic technique (right ventricular apexangiography) for positioning permanent transvenous pacemakers is described. With this technique, electrode failure occurred in only five (3%) of 186 patients followed for more than two years, as compared to an average failure rate of 23% in 11 reported series. Right ventricular apexangiography should be particularly useful in those patients in whom problems arise during insertion of a permanent transvenous electrode catheter.