Back to Search Start Over

The effect of environmental conditions and electrical charge on the weighing accuracy of different filter materials.

Authors :
Tsai CJ
Chang CT
Shih BH
Aggarwal SG
Li SN
Chein HM
Shih TS
Source :
The Science of the total environment [Sci Total Environ] 2002 Jul 03; Vol. 293 (1-3), pp. 201-6.
Publication Year :
2002

Abstract

Different filter materials and electrical charge elimination methods were used to investigate the weighing accuracy of filter papers under different environmental conditions. The results show that the standard deviations (S.D.) of weight data for glass fiber and MCE filters were substantial without environmental control, whether or not the electrical charge eliminators were used. Values of 0.157 and 0.349 mg were determined for glass and MCE filters, respectively. The accuracy of weighing was substantially improved and the S.D. was reduced to 0.01 and 0.09 mg for glass fiber and MCE filters, respectively, after applying the environmental control conditions. For PVC and Teflon filters, the accuracy of weighing was good, even in the uncontrolled environmental conditions, whether or not the electrical charge eliminators were used. The S.D. values of weighing data of PVC and Teflon filters were 0.007 and 0.011 mg, respectively.

Details

Language :
English
ISSN :
0048-9697
Volume :
293
Issue :
1-3
Database :
MEDLINE
Journal :
The Science of the total environment
Publication Type :
Academic Journal
Accession number :
12109473
Full Text :
https://doi.org/10.1016/s0048-9697(02)00015-3