Back to Search Start Over

Trace analysis of muramic acid in indoor air using an automated derivatization instrument and GC-MS(2) or GC-MS(3).

Authors :
Harley WM
Kozar MP
Fox A
Source :
Journal of microbiological methods [J Microbiol Methods] 2002 Sep; Vol. 51 (1), pp. 95-104.
Publication Year :
2002

Abstract

An automated derivatization instrument has been developed for the preparation of alditol acetates from bacterial hydrolysates for analysis by gas chromatography-mass spectrometry (GC-MS). The current report demonstrates the utility of the automated instrument for the more demanding task of trace analysis of muramic acid (Mur) in airborne dust using gas chromatography-tandem mass spectrometry (GC-MS(2)). Conditions for efficient derivatization of Mur, vital for trace analysis, are rigorous including lactam and imido group formation under anhydrous conditions. Furthermore, as the detection limit is lowered, possible contamination or carry-over of samples becomes an increasingly greater consideration and must not occur. The instrument meets these criteria and was successfully used for assaying the levels of Mur in laboratory air, which were found to be much lower than in the previous studies of heavily occupied schools and agricultural environments. The potential for GC-MS(3) in further lowering the detection limit was also demonstrated.

Details

Language :
English
ISSN :
0167-7012
Volume :
51
Issue :
1
Database :
MEDLINE
Journal :
Journal of microbiological methods
Publication Type :
Academic Journal
Accession number :
12069894
Full Text :
https://doi.org/10.1016/s0167-7012(02)00064-7