1.5-D high intensity focused ultrasound array for non-invasive prostate cancer surgery.
- Source :
-
IEEE transactions on ultrasonics, ferroelectrics, and frequency control [IEEE Trans Ultrason Ferroelectr Freq Control] 2002 Feb; Vol. 49 (2), pp. 231-42. - Publication Year :
- 2002
-
Abstract
- The aim of this study is to demonstrate the feasibility of a new spherically curved 1.5-D phased array for the treatment of localized prostatic cancer. The device is designed to conform to the Ablatherm machine (EDAP-Technomed, France), a commercially available machine in which high intensity focused ultrasound (HIFU) treatment for prostate cancer is administered transrectally. It uses high intensity electronically focused ultrasound to steer a beam along two axes, allowing enough depth to be reached to treat large prostates and eliminating two degrees of mechanical movement. Through computer simulation, it was determined that a curved 1.5-D configuration offered the optimal design. Two configurations were then proposed, and their ability to steer a beam within a target volume centered on the geometric focus of the transducer was simulated. An eight-element prototype was constructed to test the piezo-composite material and its electro-acoustical efficiency. Then, an array was constructed, and a multi-channel amplifier and control system were added, to permit remote operation. Acoustical and electrical measurements were made to verify performance. Finally, the 1.5-D array was tested in vitro on samples of pig liver to confirm the ability to induce lesions.
Details
- Language :
- English
- ISSN :
- 0885-3010
- Volume :
- 49
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- IEEE transactions on ultrasonics, ferroelectrics, and frequency control
- Publication Type :
- Academic Journal
- Accession number :
- 11885680
- Full Text :
- https://doi.org/10.1109/58.985707