Back to Search
Start Over
An optical edge detector for the accurate delineation of the light field in radiotherapy machines.
- Source :
-
Medical physics [Med Phys] 2001 Aug; Vol. 28 (8), pp. 1791-2. - Publication Year :
- 2001
-
Abstract
- A hand-held detector was developed which allows for the unambiguous and consistent determination of the edge of the light (optical) field on treatment machines conveniently and practically. The design incorporates an electronic circuit using three miniature photodiodes. The complete module is housed conveniently in a hand-held box measuring 6 cm x 10 cm x 1.5 cm. The device eliminates any observer subjectivity on the position of the edge of the light field and hence facilitates greatly any commissioning and quality assurance procedures requiring light-radiation field coincidence. Other applications are possible, including alignment of the light field on patients.
- Subjects :
- Humans
Quality Control
Light
Radiotherapy instrumentation
Radiotherapy methods
Subjects
Details
- Language :
- English
- ISSN :
- 0094-2405
- Volume :
- 28
- Issue :
- 8
- Database :
- MEDLINE
- Journal :
- Medical physics
- Publication Type :
- Academic Journal
- Accession number :
- 11548951
- Full Text :