Back to Search Start Over

Simultaneous determination of quinine and four metabolites in plasma and urine by high-performance liquid chromatography.

Authors :
Mirghani RA
Ericsson O
Cook J
Yu P
Gustafsson LL
Source :
Journal of chromatography. B, Biomedical sciences and applications [J Chromatogr B Biomed Sci Appl] 2001 Apr 15; Vol. 754 (1), pp. 57-64.
Publication Year :
2001

Abstract

The determination of quinine, (3S)-3-hydroxyquinine, 2'-quininone and (10R)- and (10S)-10,11-dihydroxydihydroquinine in plasma and urine samples is described. This is the first time the R and S configurations have been correctly assigned to the two metabolites of 10,11-dihydroxyquinine. One hundred microliter-plasma samples were protein precipitated with 200 microl cold methanol. Urine samples were 10-100 x diluted and then directly injected into the HPLC. A reversed-phase liquid chromatography system with fluorescence detection and a Zorbax Eclipse XDB phenyl column and gradient elution was used. The within and between assay coefficients of variation of the method for quinine and its metabolites in plasma and urine was less than 13%. The lower limit of quantitation was in the range of 0.024-0.081 microM.

Details

Language :
English
ISSN :
1387-2273
Volume :
754
Issue :
1
Database :
MEDLINE
Journal :
Journal of chromatography. B, Biomedical sciences and applications
Publication Type :
Academic Journal
Accession number :
11318427
Full Text :
https://doi.org/10.1016/s0378-4347(00)00577-6