A rapid screening assay for measuring urinary androsterone and etiocholanolone delta(13)C ( per thousand) values by gas chromatography/combustion/isotope ratio mass spectrometry.
- Source :
-
Rapid communications in mass spectrometry : RCM [Rapid Commun Mass Spectrom] 2000; Vol. 14 (23), pp. 2294-9. - Publication Year :
- 2000
-
Abstract
- A gas chromatography/combustion/isotope ratio mass spectrometry (GC/C/IRMS) method is described and validated for measurement of delta(13)C values of the acetate derivatives of urinary etiocholanolone and androsterone. The analysis was performed with only 2 mL of urine. The sample preparation consisted of deconjugation with beta-glucuronidase, solid phase extraction, and derivatization with acetic anhydride and pyridine. The within-assay precision of two quality control (QC) urine samples ranged from 0.5 to 2.1 CV%. The between-assay precision in the same QC urines ranged from 1.7 to 3.4 CV%. Administration of testosterone enanthate to a subject resulted in a 6 per thousand decrease in delta(13)C values from -25 per thousand (baseline) to -31 per thousand. Two weeks after testosterone administration was discontinued, the delta(13)C values remained abnormally low while the urine testosterone/epitestosterone (T/E) ratio returned to less than 6. This relatively simple method is useful for rapidly screening a large number of urine samples, including those with T/E <6.<br /> (Copyright 2000 John Wiley & Sons, Ltd.)
Details
- Language :
- English
- ISSN :
- 0951-4198
- Volume :
- 14
- Issue :
- 23
- Database :
- MEDLINE
- Journal :
- Rapid communications in mass spectrometry : RCM
- Publication Type :
- Academic Journal
- Accession number :
- 11114040
- Full Text :
- https://doi.org/10.1002/1097-0231(20001215)14:23<2294::AID-RCM165>3.0.CO;2-L