Back to Search
Start Over
An IBM PC-based system for experimental haemodynamic studies.
- Source :
-
Medical engineering & physics [Med Eng Phys] 2000 Jul; Vol. 22 (6), pp. 419-24. - Publication Year :
- 2000
-
Abstract
- An IBM PC-based real-time data acquisition, monitoring and analysis system for experimental haemodynamic studies was developed. Comprehensive haemodynamic signals, such as aortic and left ventricular pressures, aortic and coronary blood flows, two segmental lengths, two segmental thicknesses, electrocardiogram and airway pressure, were acquired and monitored to assess cardiac function. The system performs computer-aided analysis and derivations on a number of haemodynamic parameters and cardiac function indices. The system has been tested and validated extensively over a number of series of experimental haemodynamic studies to investigate the effects of anaesthetic agents, cardiovascular drugs, and changes in loading on normal and critically ischaemic myocardium of anaesthetised laboratory subjects. Without this specialised and automated system, the analysis of the data acquired from the haemodynamic studies would be too time-consuming and could not be fully performed.
- Subjects :
- Blood Pressure physiology
Diastole physiology
Electrocardiography instrumentation
Humans
Myocardial Ischemia complications
Myocardial Reperfusion Injury etiology
Software Validation
Stroke Volume
Ventricular Function, Left physiology
Hemodynamics physiology
Monitoring, Physiologic instrumentation
Myocardial Reperfusion Injury diagnosis
Signal Processing, Computer-Assisted instrumentation
Subjects
Details
- Language :
- English
- ISSN :
- 1350-4533
- Volume :
- 22
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Medical engineering & physics
- Publication Type :
- Academic Journal
- Accession number :
- 11086253
- Full Text :
- https://doi.org/10.1016/s1350-4533(00)00049-7