Back to Search Start Over

A remote flow cell for UV absorbance detection with capillary HPLC based on a single strand fiber optic.

Authors :
Wang H
Yi EC
Ibarra CA
Hackett M
Source :
The Analyst [Analyst] 2000 Jun; Vol. 125 (6), pp. 1061-4.
Publication Year :
2000

Abstract

A remote flow cell based on a single strand of fused-silica fiber optic was built for UV absorbance detection with a packed capillary HPLC system, using commercially available pumps, detection electronics (Shimadzu) and fittings. This 'off-column' flow cell design is applicable to both pressure and electro-osmotically driven systems. The goals were to minimize the linearity and light leakage problems that often limit the performance of UV absorbance detection with capillary chromatography. A linear dynamic range of 10(3) (reserpine, lambda = 220 nm), and a concentration detection limit of 5.1 x 10(-8) mol l-1 were observed. Baseline noise was measured at 3.5 x 10(-5) absorbance units (AU), with a standard deviation of 1.7 x 10(-5) AU. The illuminated volume of approximately 3 nl was optimized for capillaries with inner diameters in the range 50-100 microns, and flow rates from 100 nl min-1 to 1 microliter min-1. These modifications of readily available instrumentation have allowed the construction of a practical system for fractionating complex mixtures of peptides in small amounts, prior to mass spectrometry or additional wet chemistry steps.

Details

Language :
English
ISSN :
0003-2654
Volume :
125
Issue :
6
Database :
MEDLINE
Journal :
The Analyst
Publication Type :
Academic Journal
Accession number :
10932851
Full Text :