Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing.

Authors :
Tang Z
Lee JH
Louie RF
Kost GJ
Source :
Archives of pathology & laboratory medicine [Arch Pathol Lab Med] 2000 Aug; Vol. 124 (8), pp. 1135-40.
Publication Year :
2000

Abstract

Objectives: To determine the effects of low, normal, and high hematocrit levels on glucose meter measurements and to assess the clinical risks of hematocrit errors.<br />Design: Changes in glucose measurements between low and high hematocrit levels were calculated to determine hematocrit effects. The differences between glucose measured with meters and with a plasma glucose method (YSI 2300) also were compared.<br />Setting: Six hand-held glucose meters were assessed in vitro at low (19.1%), normal (38.5%), and high (58.3%) hematocrit levels, and at 6 glucose concentrations ranging from 2.06 mmol/L (37.1 mg/dL) to 30.24 mmol/L (544.7 mg/dL).<br />Results: Most systems, regardless of the reference to which they were calibrated, demonstrated positive bias at lower hematocrit levels and negative bias at higher hematocrit levels. Low, normal, and high hematocrit levels progressively lowered Precision G and Precision QID glucose measurements. Hematocrit effects on the other systems were more dependent on the glucose concentration. Overall, Accu-Chek Comfort Curve showed the least sensitivity to hematocrit changes, except at the lowest glucose concentration.<br />Conclusions: We strongly recommend that clinical professionals choose glucose systems carefully and interpret glucose measurements with extreme caution when the patient's hematocrit value changes, particularly if there is a simultaneous change in glucose level.

Details

Language :
English
ISSN :
0003-9985
Volume :
124
Issue :
8
Database :
MEDLINE
Journal :
Archives of pathology & laboratory medicine
Publication Type :
Academic Journal
Accession number :
10923072
Full Text :
https://doi.org/10.5858/2000-124-1135-EODHLO