Back to Search Start Over

Fully automated determination of vitamins A and E in milk-based products using a robotic station-HPLC arrangement.

Authors :
Gámiz-Gracia L
Velasco-Arjona A
Luque de Castro MD
Source :
The Analyst [Analyst] 1999 May; Vol. 124 (5), pp. 801-4.
Publication Year :
1999

Abstract

An automated method, which includes an extraction procedure based on the AOAC official method of analysis for vitamins A and E in milk-based infant formula, was developed. The whole procedure is carried out automatically in a robotic station coupled to a liquid chromatograph with a UV detector for individual separation and quantification of the vitamins. Different commercial products (milk, infant formula and nutritional supplements) were analysed, and the data were compared with those obtained using the manual AOAC extraction method. Relative standard deviations using the automated method were 5.92 and 2.57% for vitamin A and E, respectively.

Details

Language :
English
ISSN :
0003-2654
Volume :
124
Issue :
5
Database :
MEDLINE
Journal :
The Analyst
Publication Type :
Academic Journal
Accession number :
10616742
Full Text :
https://doi.org/10.1039/a900922a