Back to Search Start Over

Air emboli with Haemaccel(R)

Authors :
Gray AJ
Glover P
Source :
Anaesthesia [Anaesthesia] 1999 Aug; Vol. 54 (8), pp. 790-2.
Publication Year :
1999

Abstract

We report two cases of venous air embolism which occurred in association with infusion of Haemaccel(R) using a pressure bag. As a result of these incidents, we performed a study that showed that up to 45 ml of air can be infused into a patient from a pressurised Haemaccel(R) plastic bottle using a standard administration set. We also demonstrated that the volume of air infused was influenced by the type and size of the pressure bag and the warming of the Haemaccel(R) plastic container.

Details

Language :
English
ISSN :
0003-2409
Volume :
54
Issue :
8
Database :
MEDLINE
Journal :
Anaesthesia
Publication Type :
Academic Journal
Accession number :
10460533
Full Text :
https://doi.org/10.1046/j.1365-2044.1999.00896.x