Back to Search Start Over

A collaborative study to establish the 2nd International Standard for Antithrombin Concentrate.

Authors :
Gray E
Walker AD
Heath AB
Source :
Thrombosis and haemostasis [Thromb Haemost] 1999 Jul; Vol. 82 (1), pp. 46-50.
Publication Year :
1999

Abstract

Eighteen laboratories participated in a collaborative study to calibrate a replacement for the 1st International Standard for Antithrombin. Concentrate (88/548). Excellent agreement between laboratories, as indicated by mean % gcv of 3.3 and 5.9 for functional and antigenic assays, was observed when the candidate concentrate (96/520) was assayed against the 1st International Standard for Antithrombin, Concentrate (88/548). The functional potency was found to be 7.9% (p<0.05) lower than the antigenic potency. Based on the results and with the agreement of the participants of this study and the authorisation of the Expert Committee on Biological Standardisation of the World Health Organisation, the antithrombin concentrate, coded 96/520, has been established as the 2nd International Standard for Antithrombin, Concentrate, with labelled potencies for both functional (4.7 IU/ampoule) and antigenic (5.1 IU/ampoule) activities.

Details

Language :
English
ISSN :
0340-6245
Volume :
82
Issue :
1
Database :
MEDLINE
Journal :
Thrombosis and haemostasis
Publication Type :
Academic Journal
Accession number :
10456453