Back to Search Start Over

Relative 99mTc-MAG3 renal uptake: reproducibility and accuracy.

Authors :
Piepsz A
Tondeur M
Ham H
Source :
Journal of nuclear medicine : official publication, Society of Nuclear Medicine [J Nucl Med] 1999 Jun; Vol. 40 (6), pp. 972-6.
Publication Year :
1999

Abstract

Unlabelled: The aim of this study was to estimate the reproducibility and accuracy of 99mTc-mercaptoacetyltriglycine (MAG3) relative percentage uptake.<br />Methods: Reproducibility was evaluated on healthy volunteers who were submitted twice to a 99mTc-MAG3 renographic study, which used different uptake algorithms, different background corrections and different time intervals. Accuracy was evaluated in a group of patients with symmetrical or asymmetrical relative renal function, who underwent both 99mTc-dimercaptosuccinic acid (DMSA) and 99mTc-MAG3 studies, using the DMSA relative percentage uptake as a reference.<br />Results and Conclusion: The methods that combined the best reproducibility and accuracy for estimating 99mTc-MAG3 left-to-right uptake ratio were the integral method, with subrenal or perirenal background correction, and the Patlak-Rutland plot. The use of the integral method without background correction introduced a systematic bias, whereas the slope method resulted in high variability. Therefore these methods cannot be recommended.