Back to Search
Start Over
Real-time control of antibody loading during protein A affinity chromatography using an on-line assay.
- Source :
-
Journal of chromatography. A [J Chromatogr A] 1999 Jul 16; Vol. 849 (1), pp. 191-6. - Publication Year :
- 1999
-
Abstract
- We show that an on-line chromatographic assay can reliably control antibody loading in real-time during protein A affinity chromatography purification of a recombinant antibody from clarified Chinese hamster ovary cell culture fluid. The on-line assay directly sampled preparative column effluent and provided real-time measurement of antibody breakthrough during loading. The on-line assay used protein A immobilized on perfusion chromatography media, equilibrated with phosphate-buffered saline at pH 7.2 and eluted with phosphate-buffered saline at pH 2.2. The assay reliably ended loading at 1% breakthrough with minimal yield loss. Reproducible yield and purity were obtained over 23 consecutive cycles. Yield remained constant while breakthrough capacity and the antibody concentration in the load changed.
Details
- Language :
- English
- ISSN :
- 0021-9673
- Volume :
- 849
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Journal of chromatography. A
- Publication Type :
- Academic Journal
- Accession number :
- 10444843
- Full Text :
- https://doi.org/10.1016/s0021-9673(99)00539-7