Back to Search Start Over

Influence of latex glove hydration on bacteriophage penetration.

Authors :
Rodeheaver GT
Drake DB
Neal JG
Jackson EM
Suber F
Foresman PA
Edlich RF
Source :
Journal of long-term effects of medical implants [J Long Term Eff Med Implants] 1998; Vol. 8 (3-4), pp. 241-8.
Publication Year :
1998

Abstract

The purpose of this study was to determine whether glove hydration influenced bacteriophage penetration. Using an electronic glove hole-detection device, one brand of latex glove was identified that hydrated rapidly (3.25 min +/- 0.71 min), while another brand was selected that resisted hydration (120 min +/- 0 min). Using a standard bacteriophage penetration model, the amount of bacteriophage penetration in both the rapidly hydrating gloves and the gloves that resisted hydration was extremely small and did not differ significantly from each other.

Details

Language :
English
ISSN :
1050-6934
Volume :
8
Issue :
3-4
Database :
MEDLINE
Journal :
Journal of long-term effects of medical implants
Publication Type :
Academic Journal
Accession number :
10186970