Back to Search
Start Over
Operation of a 400 MHz NMR magnet using a (RE:Rare Earth)Ba2Cu3O7−x high-temperature superconducting coil: Towards an ultra-compact super-high field NMR spectrometer operated beyond 1 GHz.
- Source :
-
Journal of Magnetic Resonance . Dec2014, Vol. 249, p38-48. 11p. - Publication Year :
- 2014
-
Abstract
- High-temperature superconductors (HTS) are the key technology to achieve super-high magnetic field nuclear magnetic resonance (NMR) spectrometers with an operating frequency far beyond 1 GHz (23.5 T). (RE)Ba 2 Cu 3 O 7− x (REBCO, RE: rare earth) conductors have an advantage over Bi 2 Sr 2 Ca 2 Cu 3 O 10− x (Bi-2223) and Bi 2 Sr 2 CaCu 2 O 8− x (Bi-2212) conductors in that they have very high tensile strengths and tolerate strong electromagnetic hoop stress, thereby having the potential to act as an ultra-compact super-high field NMR magnet. As a first step, we developed the world’s first NMR magnet comprising an inner REBCO coil and outer low-temperature superconducting (LTS) coils. The magnet was successfully charged without degradation and mainly operated at 400 MHz (9.39 T). Technical problems for the NMR magnet due to screening current in the REBCO coil were clarified and solved as follows: (i) A remarkable temporal drift of the central magnetic field was suppressed by a current sweep reversal method utilizing ∼10% of the peak current. (ii) A Z2 field error harmonic of the main coil cannot be compensated by an outer correction coil and therefore an additional ferromagnetic shim was used. (iii) Large tesseral harmonics emerged that could not be corrected by cryoshim coils. Due to those harmonics, the resolution and sensitivity of NMR spectra are ten-fold lower than those for a conventional LTS NMR magnet. As a result, a HSQC spectrum could be achieved for a protein sample, while a NOESY spectrum could not be obtained. An ultra-compact 1.2 GHz NMR magnet could be realized if we effectively take advantage of REBCO conductors, although this will require further research to suppress the effect of the screening current. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 10907807
- Volume :
- 249
- Database :
- Academic Search Index
- Journal :
- Journal of Magnetic Resonance
- Publication Type :
- Academic Journal
- Accession number :
- 99737443
- Full Text :
- https://doi.org/10.1016/j.jmr.2014.10.006