Back to Search Start Over

LYSO crystal calorimeter readout with silicon photomultipliers.

Authors :
Berra, A.
Bonvicini, V.
Cecchi, C.
Germani, S.
Guffanti, D.
Lietti, D.
Lubrano, P.
Manoni, E.
Prest, M.
Rossi, A.
Vallazza, E.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Nov2014, Vol. 763, p248-254. 7p.
Publication Year :
2014

Abstract

Large area Silicon PhotoMultipliers (SiPMs) are the new frontier of the development of readout systems for scintillating detectors. A SiPM consists of a matrix of parallel-connected silicon micropixels operating in limited Geiger–Muller avalanche mode, and thus working as independent photon counters with a very high gain (~10 6 ). This contribution presents the performance in terms of linearity and energy resolution of an electromagnetic homogeneous calorimeter composed of 9 ~ 18 X 0 LYSO crystals. The crystals were readout by 36 4×4 mm 2 SiPMs (4 for each crystal) produced by FBK-irst. This calorimeter was tested at the Beam Test Facility at the INFN laboratories in Frascati with a single- and multi-particle electron beam in the 100–500 MeV energy range. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01689002
Volume :
763
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
98141968
Full Text :
https://doi.org/10.1016/j.nima.2014.06.012