Soft-x-ray camera for internal shape and current-density measurements on a noncircular tokamak.

Authors :
Fonck, R. J.
Jaehnig, K. P.
Powell, E. T.
Reusch, M.
Roney, P.
Simon, M. P.
Source :
Review of Scientific Instruments. Aug1988, Vol. 59 Issue 8, p1831. 3p.
Publication Year :
1988

Abstract

A two-dimensional, tangentially viewing, soft-x-ray pinhole camera has been fabricated to provide internal shape measurements on the PBX-M tokamak. It consists of a scintillator at the focal plane of a foil-filtered pinhole camera, which is, in turn, fiber-optically coupled to an intensified framing video camera (Δ t ≥ 3 ms). Automated data acquisition is performed on a stand-alone image processing system, and data archiving and retrieval take place on an optical disk video recorder. The derivation of the poloidal emission distribution from the measured image is done by fitting to model profiles. [ABSTRACT FROM AUTHOR]

Subjects

Subjects :
*TOKAMAKS
*SCINTILLATORS

Details

Language :
English
ISSN :
00346748
Volume :
59
Issue :
8
Database :
Academic Search Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
9781497
Full Text :