Back to Search
Start Over
Electron energy distribution function by using probe method in electron cyclotron resonance multicharged ion source.
- Source :
-
Review of Scientific Instruments . 2014, Vol. 85 Issue 2, p1-3. 3p. 2 Black and White Photographs, 2 Charts, 1 Graph. - Publication Year :
- 2014
-
Abstract
- We are constructing a tandem type electron cyclotron resonance (ECR) ion source (ECRIS). High-energy electrons in ECRIS plasma affect electron energy distribution and generate multi-charged ion. In this study, we measure electron energy distribution function (EEDF) of low energy region (≤100 eV) in ECRIS plasma at extremely low pressures (10-3-10-5 Pa) by using cylindrical Langmuir probe. From the result, it is found that the EEDF correlates with the electron density and the temperature from the conventional probe analysis. In addition, we confirm that the tail of EEDF spreads to high energy region as the pressure rises and that there are electrons with high energy in ECR multicharged ion source plasma. The effective temperature estimated from the experimentally obtained EEDF is larger than the electron temperature obtained from the conventional method. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00346748
- Volume :
- 85
- Issue :
- 2
- Database :
- Academic Search Index
- Journal :
- Review of Scientific Instruments
- Publication Type :
- Academic Journal
- Accession number :
- 97144047
- Full Text :
- https://doi.org/10.1063/1.4829738