Permanent Leadless Cardiac Pacing Results of the LEADLESS Trial.
- Source :
-
Circulation . 4/8/2014, Vol. 129 Issue 14, p1466-1471. 6p. - Publication Year :
- 2014
-
Abstract
- Background--Conventional cardiac pacemakers are associated with several potential short- and long-term complications related to either the transvenous lead or subcutaneous pulse generator. We tested the safety and clinical performance of a novel, completely self-contained leadless cardiac pacemaker. Methods and Results--The primary safety end point was freedom from complications at 90 days. Secondary performance end points included implant success rate, implant time, and measures of device performance (pacing/sensing thresholds and rate-responsive performance). The mean age of the patient cohort (n=33) was 77±8 years, and 67% of the patients were male (n=22/33). The most common indication for cardiac pacing was permanent atrial fibrillation with atrioventricular block (n=22, 67%). The implant success rate was 97% (n=32). Five patients (15%) required the use of > 1 leadless cardiac pacemaker during the procedure. One patient developed right ventricular perforation and cardiac tamponade during the implant procedure, and eventually died as the result of a stroke. The overall complication-free rate was 94% (31/33). After 3 months of follow-up, the measures of pacing performance (sensing, impedance, and pacing threshold) either improved or were stable within the accepted range. Conclusions--In a prospective nonrandomized study, a completely self-contained, single-chamber leadless cardiac pacemaker has shown to be safe and feasible. The absence of a transvenous lead and subcutaneous pulse generator could represent a paradigm shift in cardiac pacing. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00097322
- Volume :
- 129
- Issue :
- 14
- Database :
- Academic Search Index
- Journal :
- Circulation
- Publication Type :
- Academic Journal
- Accession number :
- 95635342
- Full Text :
- https://doi.org/10.1161/CIRCULATIONAHA.113.006987